Sweep Trigger Source
In the sweep mode, the U2761A outputs a single sweep when a trigger signal is
received. After one sweep from the start frequency to the stop frequency, the
U2761A will continuously output the preset waveform at the start frequency while
waiting for the next trigger.
– The sweep trigger source consists of Internal (immediate), External, or Manual
– When the Internal source is selected, the U2761A outputs a continuous sweep
– When the External source is selected, the U2761A accepts a hardware trigger
– When the Manual source is selected, the U2761A outputs one sweep each
Soft Front Panel Operation
On the Trigger Setup panel as shown in
trigger source from the drop down list
To specify whether the U2761A is triggered on the rising or falling edge of the
signal from the Trig In connector, select Positive
triggering and Negative
Remote Interface Operation
Use the following command to specify whether the U2761A is triggered on the
rising or falling edge of the signal from the Trig In connector.
Keysight U2761A User's Guide
(software). The default trigger source is Internal.
at a rate determined by the specified sweep time.
applied to the device Trig In connector. The U2761A initiates one sweep each
time Trig In receives a TTL pulse with the specified polarity.
time the soft front panel Trigger
TRIGger:SOURce {IMMediate|EXTernal|BUS}
TRIGger:SLOPe {POSitive|NEGative}
Features and Functions
is clicked.
Figure
2-28, select the desired sweep
at the Source panel.
for falling edge triggering.
for rising edge
2
89