DC Offset Voltage
The default offset is 0 V for all functions.
Limits Due to Amplitude
The relation between the offset voltage and output amplitude is shown below.
Vmax is the maximum peak voltage for the selected output termination (5 V for a
50 load or 10 V for a high-impedance load).
V offset
If the specified offset voltage is not valid, the U2761A will automatically adjusts it
to the maximum DC voltage allowed with the specified amplitude.
Soft Front Panel Operation
On the Offset panel as shown in
select the unit from the drop down list.
Figure 2-5
Remote Interface Operation
Or, you can set the offset by specifying a high level and low level using the
following commands.
You can also use the APPLy command to select the function, frequency,
amplitude, and offset.
Keysight U2761A User's Guide
V pp
V max
- - - - - - - -
–
2
Panel view of the DC offset section
VOLTage:OFFSet <offset>
VOLTage:HIGH <voltage>
VOLTage:LOW <voltage>
Features and Functions
Figure
2-5, input the desired offset value and
2
41