Phase Deviation
The phase deviation setting represents the peak variation in phase of the
modulated waveform from the carrier waveform. The phase deviation setting can
be set from 0 to 360. The default phase deviation setting is 180.
Soft Front Panel Operation
On the Phase Deviation panel as shown in
deviation value.
Remote Interface Operation
Generate PM Using SCPI Commands
Example 1
-> *CLS; *RST
-> PM:STAT ON
-> FUNC SIN
-> FREQ 1000
-> VOLT 5 VPP
-> VOLT:OFFS 0
-> PM:INT:FUNC SIN
-> PM:INT:FREQ 500
-> PM:DEV 180
-> OUTP ON
Keysight U2761A User's Guide
PM:DEViation <deviation in degrees>
//To reset the Keysight U2761A to default
//Enables PM.
//Sets the carrier waveform to Sine wave.
//Sets the carrier frequency to 1 kHz.
//Sets the output amplitude to 5 Vpp.
//Sets the output offset to 0.
//Changes the modulating waveform shape
//Changes the modulating frequency to
//Changes the phase deviation to 180.
//Turns on output.
Features and Functions