Features and Functions
Sweep Mode
You can sweep with either linear or logarithmic spacing. For a linear sweep, the
U2761A varies the output frequency in a linear fashion during the sweep. For a
logarithmic sweep, the U2761A varies the output frequency in a logarithmic
fashion.
The sweep mode consists of Linear or Logarithmic. The default sweep mode is
Linear.
Soft Front Panel Operation
On the Sweep Type panel as shown in
Log
Remote Interface Operation
Sweep Time
The sweep time specifies the number of seconds required to sweep from the start
frequency to the stop frequency. The number of discrete frequency points in the
sweep is automatically calculated by the U2761A and is based on the sweep time
you have selected. The range of the sweep time is 1 ms to 500 s. The default
sweep time is 1 s.
Soft Front Panel Operation
On the Sweep Time panel as shown in
and select the unit from the drop down list.
Remote Interface Operation
88
mode.
SWEep:SPACing {LINear|LOGarithmic}
SWEep:TIME <seconds>
Figure
2-28, select the Linear
Figure
2-28, input the desired sweep time
Keysight U2761A User's Guide