Optical safety
Possibility of glare.
Electromagnetic disturbance
Incorrect use
Do not place objects on Product.
Improper use of mobile version
Pushing or resting on the product
is forbidden.
MU_023
Rev.2
MAINTENANCE MANUAL

2.2 Safety conditions (secondary effects)

Do not direct the light source into the patient's and/or operator's
-
eyes.
When Product use is restricted to the face (maxilla-facial
-
surgery, plastic surgery, ear-nose-throat surgery) the patient's
eyes must be covered with adequate protection.
Failure to follow such precautions could cause glare and
potential damage to the retina.
To avoid any significant risk of reciprocal interference due to the
presence of the Product during specific exams or treatments, refer
to section 10 of the Manual.
Never place and/or hang anything on the Product.
-
Failure to follow such precaution could result in such objects
falling in the operating area.
Never hang on the Product with the body weight of a person.
-
Failure to follow such precaution could damage the Product
structure.
Never cover the head of the Product during operation to prevent
-
overheating.
Avoid the Product parts colliding with one another or other
-
nearby equipment.
Knocks could cause the detachment of plastic parts or paint from
the Product which could fall in the patient area.
In the case of the mobile version, do not rest, push or lie on the
product. Failure to comply could result in damage to the product
and to devices nearby and injury to staff members.

2.3 Environmental conditions

The Product is not suitable for use in explosion-risk areas.
-
The Product is not suitable for use wherever there are
-
inflammable mixes of anaesthetics with air, oxygen or N
(laughing gas).
The Product is not suitable for use in environments rich in
-
oxygen and use is not intended in the presence of inflammable
agents.
During operation, the ambient temperature must be between
-
10°C and 40°C.
Relative humidity must be between 30% and 75%.
-
Atmospheric pressure must be between 700 and 1060hPa.
-