Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
MU_023
Rev.2
Test level to
IEC 60601-1-2
3 Veff
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
28/10/2019
OPERATION AND
MAINTENANCE MANUAL
Conformity level
3 Veff
3 V/m
Electromagnetic environment -
directives
Portable and mobile RF communications
equipment should be used no closer to any
part of the Products, included cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d = 1.2√P 150 KHz to 80 MHz
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.7 GHz
where P is the maximum output power
rating of the transmitter in watts (W),
according to the transmitter manufacture
and d is the recommended separation
distance in meters (m).
Field strengths from fixed transmitters, as
determined by an electromagnetic site
survey, should be less than the compliance
leave in each frequency range.
Interference may occur
in the vicinity of equipment marked with
the following symbol:
EN
Page 28 of 32