Tecno-gaz SLIM Notice D'utilisation Et D'entretien page 58

Scialytique chirurgical secondaire
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

General data
Colour
Directive
Standards
Classification of Medical Device
does not vary by more than 20% during use; the colour temperature and the colour rendering
index are stable and are within the range 3000K-6700K and 85-100, respectively; E
Essential
performance
Limitation of energy in the operating field (UV-irradiance for wavelengths below 400 nm does not
exceed 10 W/m
IP degree of protection
Operating conditions
Handpiece steam sterilization
Mains power voltage insulation means
Dimensions
Diameter of lamp body [cm]
Light emission surface [cm
Weight of single ceiling, double ceiling, mobile,
mobile battery, wall surgical light [kg]
Markings
All technical light measurements are to be deemed with a tolerance of ±6% for metrological and manufacturing
reasons
MU_023
Rev.2
Distribution of minimum and adequate lighting (luminous flux emitted by the ME equipment
2
and the total irradiance Ee in the lighted area does not exceed 1000 W/m
distance of 1000 mm; E
2
]
28/10/2019
OPERATION AND
MAINTENANCE MANUAL
93/42/CEE and 2007/47/EC
be ≥ 40,000 lux and ≤ 160,000 lux).
value shall be ≥ 40,000 lux and ≤ 160,000 lux; E
c
121°C at 1.3bar from 25 to 30 minutes.
134°C at 2.3bar for 4 minutes.
Outside the product (main switch) for ceiling versions
Main switch for mobile and wall versions
In conformity with Directive 93/42/EEC (and 2007/47/EC)
RAL 9003
IEC 60601-2-41
Class I
/E
e
IP20
Continuous operation
40
305
13, 20, 21, 24, 12
EN
value shall
c
2
at a
≤ 6 mV/m
2
lx).
c
Page 24 of 32