Télécharger Imprimer la page

CASTELLINI Puma Eli Mode D'emploi page 159

Masquer les pouces Voir aussi pour Puma Eli:

Publicité

Les langues disponibles

Les langues disponibles

ENVIRONMENTAL CONDITIONS PERMITTED FOR TRANSPORT AND WAREHOUSING
- Ambient Temperature between -20 - +70
- Relative Humidity 10 - 100 %
- Atmospheric Pressure 500 ÷ 1060 hPa (500 ÷ 1060 mBar)
TROUBLESHOOTING
ANOMALIES
Insufficient polymerisation
The curing light shuts off before
the end of the time set.
The curing light does not turn on
Scrapping
Scrapping the device must be carried out with respect for the laws in force for electrical and electronic equipment,
according to the individual national laws.
The materials used for the manufacture are not harmful for human beings or animals that come into contact with or are
exposed to them.
All manuals and user guides at all-guides.com
PROBABLE CAUSES
Fibre optic dirty or damaged
Selected time inadequate
Activation of the reduced power mode.
Device damaged.
Intervention of the safety system
(indicated by the flashing of the LEDs)
Connection hose damaged
Device damaged.
RECOMMENDED INTERVENTION
Clean or replace the fibre optic.
Verify the time selection.
Verify the operational mode.
Request technical service assistance
Stand by for the recovery of
operation.
Request technical service assistance
Request technical service assistance
31/01/2011 - 97050183-03-GB - 74