"IMPLANTOR LED", "HANDY POWER" and "HANDY POWER L" ELECTRICAL MICROMOTORS
TECHNICAL FEATURES
Power supply
Max absorption
Max Electrical power absorbed
Max speed
Min speed
Max torque
Cooling
Intermittent Operation
Pneumatic power supply
Water Supply:
Consumption:
Classification:
Handpiece Attachment
LED Illumination
Applicable standards
Dimensions
Weight
Intended Use
Medical device, or a medical device accessory for professional use in Dentistry, possibly assisted by authorised office
personnel. The device is meant for use on dental units complying with the 93/42/EEC Directive and installed in
locales with an electrical system that complies with the IEC 60364-7-710 Regulations.
IMPORTANT WARNINGS
Before use it is required that the instructions in this manual be read carefully. The use of the device must be
performed with respect for those instructions supplied. For the correct interpretation of the indications in this manual
the Italian text is binding.
The device is not suitable for use with blends of inflammable anaesthetics and oxygen or Nitrous Oxide.
The instrument is provided in a non-sterile state. Before its use see to it suitable cleaning and disinfection. If
the device may be sterilised, lubricate it before performing the autoclave sterilisation.
Use of electrically powered devices may interfere with the operation of pacemakers. In case of doubt on the
treatment of patients with implanted cardiac stimulators it is recommended that a cardiology expert be
consulted.
During use dust and fragments from the patient's oral cavity or from the same device may be projected into the
environment (organic and inorganic dust and metal dust, fragments of the device or its tips and potentially
infected biological material)
- Protect the patient, when possible, by way of the use of the dam.
- Instruct the patient to breathe with the nose in cases where the dam is not applicable.
- Medical personnel must put on suitable personal protection equipment.
Suitably cool the surgical field during use.
Use only handpieces and counter angles with the CE mark of compliance with the 93/42/EEC
Directive.
Use only burrs(or other similar tools) with the CE Mark of compliance with 93/42/EEC Directive
made with biocompatible materials in compliance with IEC 60825-1 Regulations.
Before each use, verify the proper locking of the handpiece onto the Electrical micromotor and of the burr in
the handpiece chuck.
Do not use damaged or worn handpieces.
Make sure that during operation of the motor there is full flow of the cooling air. In absence of cooling stop
using the device and call authorised technical service assistance.
In case of visible damage, the emission of unusual noises and/or vibrations or if overheating is noticed, do not
use the device and call authorised technical service assistance.
All manuals and user guides at all-guides.com
Implantor LED
50,000 ± 10% revs/min
100 ± 10% revs/min
Class IIa (Directive 93/42 CEE) and following introduced
Compliant with ISO 3964 and compatible INTRAmatic ® Lux.
32 V DC ±6V (Tolerance)
6 A
170 W
40,000 ± 10% revs/min
1000 ± 10% revs/min
4.5 Ncm
Forced Air
5 min of work - 5 min of rest
420 ± 20 kPa (4.2 ± 0.2 bar)
120 ± 20 kPa (1.2 ± 0.2 bar)
Cooling air ~ 33 Nl/min
Air Spray~ 33 Nl/min
Water Spray 35 cc/min
modifications – (ECC Directive 2007/47)
Class II type B (IEC 60601-1)
3500 K - 20000 mlm
IEC 60601-1 – IEC 60825-1
ISO 3964 – ISO 10993 - ISO 11948
47.5 mm
82 g
31/01/2011 - 97050183-03-GB - 59
Handy Power
Handy Power L
4.5 A
120 W
3 Ncm
38.5 mm
75 g