Fig. 2 – Connecting and disconnecting handpiece
Fig. 3 – Lubrificazione manipolo
CLEANING, DISINFECTING, STERILIZING
Use gauze or cotton soaked in surgical alcohol to clean and/or disinfect micromotor exterior.
DO NOT soak micromotor directly in solution.
DO NOT use an ultrasonic cleaner.
The air-driven micromotor must be sterilised in a steam autoclave at max. 135°C, 220 kPa (2.2 bar) for 5
minutes (250 cycles tested – EN ISO 11498).
In order to maintain a proper level of hygienic safety, autoclave after the use on each patient.
Never sterilize air driven micromotor in dry-heat sterilizers.
Before and after each autoclaving lubricate the air driven motor. Never leave it in autoclave overnight: always remove it
from autoclave after each cycle.
Warning!
Check the autoclave periodically according to the manufacturer's prescriptions!
Temperature exceeding the above stated limit may damage the micromotor!
All manuals and user guides at all-guides.com
CONNECTING AND DISCONNECTING HANDPIECE
Handpieces compling with ISO 3964 may be fitted.
Fit the handpiece over the connection stem and push until it
clicks. Be sure the handpiece can swivel freely. Damaged or
worn handpieces are not be used.
For disconnecting a handpiece from the air air driven
micromotor push button.
SERVICING
Regular maintenance of the filters in the unit and draining
the condensate from the compressor tank is recommended.
Use the spray-can "DAILY OIL" to lubricate at least once
daily and in the hereafter indicated circumstances the air
driven
micromotor.
Follow
remembering to apply lubricant in two short sprays.
Eliminate any excess lubricant before use by operating the
air driven micromotor and wiping in with gauze or cotton.
In addition it is imperative to lubricate as follows:
- Before and after autoclaving
- Before and after every AUTOSTERIL disinfecting or
sterilizing cycle.
Regular use of this spray lubricant assures best working
order.
Once a week lubricate the three O-rings on the handpiece
connection stem: use the specific grease S1 distributed by
CASTELLINI S.p.A. Wearing protective disposable gloves,
apply a small amount of S1 on fingers and lubricate by it.
O-rings are supplied as spare parts together with the air
driven micromotor.
Any other technical assistance to the air driven micromotor
must be carried out by specialized personnel, authorized by
CASTELLINI S.p.A. Indossando guanti monouso, applicare
un velo di S1 sui polpastrelli e lubrificare.
A pakage of O-rings is supplied with the micromotor
Should you observe a significant increase in the build up of
condensation inside the filter situated beneath the dental
unit instrument tray, remove the cover by turning it
clockwise about a ¼ of a turn, replace the filtering element
and eliminate the liquids present; the filtering element must
be replaced for the first time within a year of the date
shown on the cover and at least once a year thereafter.
31/01/2011 - 97050183-03-GB - 65