on your breathing It may take up to 20 minutes for the SimplyGo output to
stabilize within its O
from the SimplyGo at any time
Note: The Low Oxygen Concentration Alarm is inactive for approximately 20
minutes after startup The No Breath Alarm is inactive for the rst ten
minutes after startup
Note: If the device is stored at the minimum storage temperature between
uses, please allow 25 hours for the unit to adequately warm up
Note: If the device is stored at the maximum storage temperature between
uses, please allow 25 hours for the unit to adequately cool down
Operating Modes
This section describes additional operating screens on the SimplyGo device
1. Pulse Mode Screen
When the device is in the Pulse operating mode, it detects when the user
begins to take a breath and then delivers a pulsed volume of oxygen
during the inhalation period The volume of the oxygen pulse is depen-
dent on the setting value If no breath is detected for approximately 30
seconds, the system automatically delivers the pulsed volume of oxygen
determined by the setting number at a fixed rate of 12 breaths per min-
ute If no breath is detected for 2 minutes, the device signals an alarm and
shuts down after 30 minutes
2. Continuous Flow Mode Screen
In Continuous Flow operating mode, a constant flow of oxygen is sup-
plied based on the setting number (in liters per minute):
5 LPM
•
•
1 LPM
•
15 LPM
•
20 LPM
purity specications However, you may begin breathing
2
SimplyGo User Manual
17