SimplyGo User Manual
EMC Information
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specied below The user of this device should assure that it
is used in such an environment
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
Emission of Radio Frequency Energy
RTCA/DO-160G Section 21
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specied below The user of this device should assure that it
is used in such an environment
Immunity Test
IEC 60601 Test Level
Electrostatic
±8 kV contact
Discharge (ESD)
±2 kV, ±4 kV, ±8 kV, &
IEC 61000-4-2
±15 kV Air
Electrical Fast
±2 kV for Power Supply
Transient/Burst
Lines
IEC 61000-4-4
±1 kV for Input/Output
Lines
Surge
±1 kV Line to Line
IEC 61000-4-5
±2 kV Line to Ground
Voltage Dips, Short
0% U
05 cycle,
T
Interruptions and
45 deg inc
Voltage Variations on
Power Supply Input
0% U
1 cycle
T
Lines
IEC 61000-4-11
70% U
25 Cycles
T
(30 cycles if US)
0% U
5 sec
T
Power Frequency
30 A/m
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level
T
46
Compliance
Electromagnetic Environment - Guidance
Group 1
The Device uses RF energy only for its internal
function Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment
Class B
The Device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
Class A
the public low-voltage power supply network
that supplies buildings used for domestic
purposes
Complies
Category M
The Device is suitable for use onboard
commercial airplanes inside passenger cabin
Compliance Level
Electromagnetic Environment - Guidance
±8 kV contact
Floors should be wood, concrete or ceramic tile
If oors are covered with synthetic material, the
±2 kV, ±4 kV, ±8 kV, &
±15 kV Air
relative humidity should be at least 30%
±2 kV for Power Supply
Mains power quality should be that of a typical
Lines
home or hospital environment
NA- Device does not have
user I/O lines that are
longer than 3m in length
±1 kV Line to Line
Mains power quality should be that of a typical
NA- The device is a Class
home or hospital environment
II device and does not
connect to earth ground
0% U
05 cycle,
Mains power quality should be that of a typical
T
45 deg inc
home or hospital environment If the user of the
Device required continued operation during
0% U
1 cycle
power mains interruptions, it is recommended
T
that the Device be powered from an
70% U
25 Cycles
uninterruptible power supply or battery
T
(30 cycles if US)
0% U
5 sec
T
Power frequency magnetic elds should be at
30 A/m
levels characteristic of a typical location in a
typical home or hospital environment