Warning: It is very important to set your device to your prescribed level of
oxygen ow Do not increase or decrease your ow rate from your
prescribed level until you rst consult with your physician or your
healthcare professional
5
Position the nasal cannula on your face, as shown below, and breathe
normally through your nose
To verify the alarm system:
1
Ensure the device is in Continuous mode Induce a No Flow Alarm
condition by turning the device ON and completely blocking the patient
cannula connector of the device
2
Wait approximately 45 seconds or until the No Flow Alarm occurs
3
Verify that the No Flow Alarm indicators and screen symbols are present
4
Unblock the patient cannula connector for a minimum of 45 seconds
5
Verify that the No Flow Alarm ceases
Turning the SimplyGo Device O
1
When you are ready to turn your SimplyGo device off, press the Power
button The screen below appears
2
Press the Power button a second time and the SimplyGo device will turn
off
3
Rather than pressing the Power button a second time, the backlight
intensity may be increased or decreased by repeatedly pressing the Plus
(+) or Minus (-) key until the desired backlight intensity is achieved The
device may then be powered off using steps 1 and 2 above
SimplyGo User Manual
21