2.
Tighten or loosen the main screw (10) using a 4 mm Allen key.
3.
Lock the central screw (9): tightening torque 2.5 Nm.
F. Version change
Switching from 1M01 to 1M10:
Remove the two screws (11), taking care not to lose the joint axes.
Select position A or B for the handle (12).
Mount the clevises (13) on the handle (12).
Engage the ends of the handle (12) into the lever holes (2).
Secure the assembly using the screws (14).
Switching from 1M10 to 1M01:
Remove the screws (14), clevises (13) and handle (12).
Install the cable as shown above.
In place of the screws (14), refit the hex headless screws contained in the
cable's package
G. Finishing
The socket must be in contact with the tube holding collar at
maximum bend. Any other contact zone is prohibited.
H. Installing the 1M01100 unlocking system
1.
Insert the tapped insert (1) into the wood or laminate.
2.
Assemble the washer (2), handle (3) and ring (4) with the screw (5).
3.
Cut the sheath to the desired size.
4.
Insert the cable into the hole in the handle (3), and into the sheath, then fasten it on the
lock lever with the desired tension.
5.
The handle will be angled approximately 30° above horizontal in the locked position.
6.
Fit one or two bridge fittings (8) using screws or tubular rivets (7).
7.
Possibly fasten the release holding tab (6) using screws or tubular rivets (7).
7. DETECTING MALFUNCTIONS
If you observe abnormal behaviour or feel modifications in the device's properties, or if it undergoes
a heavy impact, please consult your orthopaedic prosthetist.
8. WARNINGS, CONTRAINDICATIONS, SIDE EFFECTS
A. Warnings
To avoid degrading the knee, do not use talc to eliminate friction noise, but a silicone spray
instead. Talc causes deterioration of the mechanical components, which can lead to a malfunction with a risk of the patient falling.
PROTEOR disclaims all liability if the knee is used with talc.
There is a risk of the knee joint trapping a finger or catching clothing. To prevent any risk of injury due to joint movement, make
sure that no-one puts their fingers near or inside the mechanism.
The knee can withstand bad weather, but needs to be dried after being soaked.
B. Contraindications
It is strictly prohibited to tighten or loosen any screw on this knee, with the exception of the three setting screws 1, 9, 10 and
tightening screws V1 (+ V2), which are intended for use by the orthopaedic prosthetist.
Never grease the knee axes, as this could quickly degrade them.
The warranty excludes deterioration caused by misuse, unsuitable alignment, use in a very dusty environment and without
suitable protection, or any inappropriate use.
Avoid exposing the knee to environments that could cause corrosion of the metal parts (soft water, seawater, chlorinated water,
acids, etc.).