3.5.1.
Inspecting parts
MN030-616-S1 EN
A
Jaw
A1
Jaw OPEN/CLOSE
A2
Check straightness.
A3
Check condition of ball
(distortion, fracture, or similar).
Check jaw function.
Discard or replace defective
devices.
B
Shaft tube
Check Check the insulation
visually for possible
damage.
Discard or replace
defective devices.
C
Handle
Install C1
Check:
•
Do C1, C2, C3 move
freely?
Discard or replace
defective devices.
D
HF cable
Check insulation visually.
Dispose of HF cables with
worn or faulty insulation.
13