2.
On delivery
Target group
Only trained medical personnel in medical establishments.
Delivered condition
The product is originally packed by BOWA (delivery unsterile).
Reading the Instruction Manual
 Before use, read this Instruction Manual thoroughly.
 Read the instruction manuals of other products used during the procedure
thoroughly.
Storing this Instruction Manual
 Store in a safe, easily accessible place.
The storage location must be known.
Storage and transportation
Storage, page 17
Before use
WARNING
Use only sterilised products in perfect condition.
 Inspect and decontaminate product.
3.
Application
What you should know when connecting the bipolar instrument
3.1.
Preparation
Applying the return
plate
MN030-616-S1 EN
 Preparation, page 15
 Attach return plate close to operation area, if
possible on upper arm or thigh.
 Render the skin at this location as far as possible
free from hair and fats.
 Ensure contact over the entire area.