Active electrodes
HF unit
Endogenous burns
Exogenous burns
MN030-616-S1 EN
WARNING
Risk of unintended burns.
 Never use worn out or faulty active
electrodes.
 Dispose of worn out or faulty active
electrodes - never repair!
Pay attention to the electrosurgical generator
instruction manual and to general information
on electrosurgical procedures.
 Technical data, page 6
WARNING
Improper use of HF current can cause endogenous and
exogenous burns as well as explosions.
 Carry out electrosurgical procedures only with
insufflation of incombustible gases (CO
Endogenous burns are burns caused by high local
current density in the tissue of the patient.
Causes can include:
With direct skin contact by HF cables, capacitive
effects can lead to burns.
Exogenous burns are burns caused by the heat of
ignited liquids or gases. Even explosions are possible,
through detonation.
Causes can include:
Ignition of skin cleaning solution and disinfectant.
Ignition of anaesthetic gases.
Ignition of insufflated gases.
 Only insufflate incombustible gases!
Ignition of endogenous gases (e.g. in the intestine).
).
2