Procedure
 Position the jaw at the operating field.
 Close jaw.
 Activate HF current for coagulation.
 Open jaw.
 Remove jaw from the coagulated tissue (The tips may still be so hot just after
activation that it can cause burns).
Extraction
 Close jaw.
 Remove monopolar instrument from the trocar sleeve.
WARNING
It cannot be excluded that a movable part of the jaw will be damaged
through excessive force and where possible even broken off.
 After every procedure, inspect the jaw. All parts must be present.
 Inspecting products, page 12
3.4.
After use
 Decontaminate.
 Preparation, page 7
3.5.
Inspecting products
WARNING
During use as intended, these products are subject to lighter or
heavier wear depending on intensity of use.
This wear has technical causes and is unavoidable.
 Inspect and decontaminate product.
 If the product reveals externally-recognisable faults or does not function as
described in this Instruction Manual, without delay advise the manufacturer or its
authorised representatives.
 Replace faulty products.
12
MN030-616-S1 EN