5.1 Cleaning overview
Washing and sanitizing are two different activities. They must be done separately to protect
you and your baby from harm, and to maintain the performance of the device.
Wash – To clean the surfaces of the parts by physically removing contamination.
Sanitize – To kill microorganisms, such as fungi, bacteria or viruses, that may be present on
the surfaces of the parts.
When to clean
Before first use
After each use
Once per day
When to clean
As needed
101043399-2021-04-A-swing-maxi-MedelaFamily-MD-ifu-ENUS-FRCA-ESMX.indb 13
101043399-2021-04-A-swing-maxi-MedelaFamily-MD-ifu-ENUS-FRCA-ESMX.indb 13
Breast
Connector bodies
shields
and membranes
Disassemble the parts according to section 5.2
Wash the parts according to section 5.3
Sanitize the parts according to section 5.4
Disassemble the parts according to section 5.2
Wash the parts according to section 5.3
Disassemble the parts according to section 5.2
Sanitize the disassembled, washed parts according to section 5.4
Tubing
Normally, cleaning of the tubing
is not necessary.
Refer to section 5.6.
Bottles & lids
Pump unit
swing maxi
Wipe with clean, damp towel.
Refer to section 5.5.