Chapter 4 - Technical Specification
Feature description
Power Consumption
The touch screen
be powered through the
MUST
scan head and is included in these figures.
Touch screen voltage
4.1.2 Connections
These connections are
NOT
1
M13385001
2
3
FIGURE 10: Sample of the connection label
No. Description
Mains power cable and switch
1
Protective earth must be fitted.
l
U.K. Power Cable Fuse: 13A
l
2
Network cable
3
Hand control and Optos Service use
4
Touch screen
4.1.3 Scan head dimensions
The scan head, hand control and touch screen should be placed on a table which meets the table specification,
see Table requirements on the facing page . There should be adequate clearance around the table to operate
the device.
The image server, switch and any browser-based review clients or monitors must be positioned outside the
patient environment; more than 1.5 meters (5 ft) from the scan head.
Page 56 of 69
English
Compliance
300VA
3A at 100-120Vac
1.5A at 200-240Vac
SMPS continuous
+12V
intended to be detached by the user.
100-240V AC
T5A H250V
4
Part Number: G108025/4GWE
Copyright 2017, Optos plc. All rights reserved.