3.2.4 How to capture SLO images

To capture SLO images use the procedure below. You have the option to capture the following image
modalities:
opto map
l
opto map
plus
l
opto map
af
l

FIGURE 8: SLO capture workflow

1. Select the required SLO light and resolution types. You should select
or opto map
plus
2. Position the patient, see How to prepare for image capture on page 36 .
3. Align the patient, see How to align the patient on page 37 .
4. Ask the patient to open both eyes wide, look at the center of the alignment target and to move closer
until they can see the whole target.
5. Capture the image either manually or using auto-capture.
Auto-capture : Auto-capture is enabled when it is set to
l
captures an image when the patient is correctly aligned. You can press the screen to
manually capture the image.
Manual capture : Press the screen to capture the image when the pupil is in the center of
l
the cross hairs and the patient alignment target is green.
6. Check the quality of the images, see What to do after the image is captured on the next page .
Part Number: G108025/4GWE
Copyright 2017, Optos plc. All rights reserved.
.
Chapter 3 - How to...
Red/Green
AF
or
and opto map
On
on the touch screen. The system
Page 39 of 69