KaWe MASTERLIGHT LED 1000 Mode D'emploi page 38

Masquer les pouces Voir aussi pour MASTERLIGHT LED 1000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
Table 3: AC 2-pole for power input (according to IEC 60601-1-2).
Table 3: AC 2-pole for power input (according to IEC 60601-1-2).
Phenomenon
ment
Electrical fast transient/
impulse withstand voltage a) l) o)
impulse voltages
impulse voltages
Conducted interference caused by high
frequency fields
al plane of
Voltage dips
) p) r)
t the
Voltage interruption
he basic
r)
during
a) The test can be carried out at any mains input voltage in the range of RATED VALUES for ME-DEVICE or ME-SYSTEM mains voltage. If ME-DEVICE or ME-SYSTEM are
tested at the mains input voltage, additional tests with other voltages are not required.
b) During the testing, all the wires of ME-DEVICES and ME-SYSTEMS must be mounted.
c) The calibration of the current supply clamps must be done on the 150.Q system.
he test
d) If the frequency steps are outside the ISM or amateur radio band, an additional test frequency must be used within the ISM or amateur radio band. This applies to
all ISM or amateur radio bands within the specified frequency range.
e) The test may also be conducted at other baseband frequencies as determined by the RISK MANAGEMENT PROCESS.
f) ME-DEVICES and ME-SYSTEMS with a DC power input, designed for operation with a transformer with an AC input and a DC output, must be tested with a
k analysis
transformer corresponding to the parameters specified by the MANUFACTURER of the ME-DEVICE or ME-SYSTEM. IMMUNITY TEST LEVELS must be applied to the
unity
transformer AC input.
g) Applies only to ME-DEVICES and ME-SYSTEMS designed for connection to single-phase AC mains.
h) Slash-separated data, for example 10/12, means 10 cycles at 50 Hz or 12 cycles at 60 Hz line frequency.
38
EMC basic standard
IEC 61000-4-4
a) b) j) k) o)
line-to-line
IEC 61000-4-5
a) b) j) k) o)
line-to-earth
IEC 61000-4-5
IEC61000-4-6
c) d) o)
IEC 61000-4-11
f)
'
o)
IEC 61000-4-11
Immunity test level
Professional health care institutions
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2 kV
3 V
m)
0.15 MHz - 80 MHz
6 V
m)
in ISM bands from 0.15 MHz to 80 MHz
n)
80% AM at 1 kHz
e)
0% U
; for 0.5 cycle
g)
T
at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% U
; for 1 cycle and
T
70% U
; for 25/30 cycle
h)
Single phase: at 0
T
degrees
0% U
; for 25/30 cycles
h)
T
Home health care environment
3 V
m)
0.15 MHz - 80 MHz
6 V
m)
in ISM and amateur radio bands from
0.15 MHz to 80 MHz
n)
80% AM at 1 kHz
c)
q)

Publicité

Table des Matières
loading

Table des Matières