Aesculap
Sterile Container System Edition 01/2013
5.4
Manual cleaning/disinfecting
►
Prior to manual disinfecting, allow water to drip off for a sufficient
length of time to prevent dilution of the disinfecting solution.
►
After manual cleaning/disinfection, check visible surfaces visually for
residues.
►
Repeat the cleaning/disinfection process if necessary.
Manual cleaning with disinfecting cleaning by wipe disinfection
Phase
Step
I
Cleaning
II
Dry
III
Wipe disinfection
IV
Final rinse
V
Dry
D–W:
Drinking water
FD–W:
Demineralized water
RT:
Room temperature
Phase I
►
Wipe off gross debris prior to cleaning.
Phase II
►
Dry the instrument with a lint-free cloth or medical compressed air.
Phase III
►
Carry out wipe disinfection
Phase IV
►
When using aldehyde-free disinfectants:
– After the specified exposure time (at least 1 min), rinse the disin-
fected surfaces under running FD water.
►
Drain any remaining water fully.
Phase V
►
Dry the product with a lint-free cloth or in a heating cabinet.
8
®
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) -
-
RT
-
-
-
>1
-
RT (cold) 0,5
-
RT
-
-
Water quality
Chemicals
D–W
-
-
-
-
I
Alcohol Denat. 70 %
(Ethanol B|BRAUN)
FD–W
not required
-
-
II
Aldehyde-free surface disin-
fectant (Stabimed)
Rinse off cleaning chemicals;
do not leave residues