OPERATION
The hose connecting the compressor to medical ventilator must not
pass through a cold environment i.e. placed on the ground. It should be
as short as possible with no kinks (this may cause water to condense
inside the hose).
The compressor must not be used for supplying air to the medical air
central distribution pipeline system.
Electrical connection
The compressor comes with a plug containing an appropriate
protective contact (ground) adhere to local electrical regulations. The
voltage
specifications on the data label.
The electrical cord must not be stressed or have any tension exerted
upon it, and must always be free.
The socket must be accessible for safety reasons so that the equipment
can be safely disconnected from the power supply in case of an
emergency.
The relevant current circuit must be protected.
First operation
Make sure that the pump inside of device was unsecured.
Check that the connection to the compressed air supply is correct.
Check for proper connection to the main power supply. Before the
equipment is plugged in, make sure that the voltage and frequency shown
on the equipment are the same as provided by the power connection.
Switch on the pressure switch to position "I".
OPERATION
The equipment may only be operated by the trained staff!
In case of emergency, switch the equipment off at the switch and pull
out the main power plug.
Switching the compressor on
The compressor is switched on at the main power switch by putting it in
position "I". The compressor operates in one of the following modes
depending on the pressure in the central distribution line and on the
consumption of compressed air:
and frequency of the
mains
must comply with the