Installation of couplings for connecting hoses
Use a box wrench to install the couplings supplied with the product into the
openings in the rear panel. Beware of possible confusion - the marking on
the coupling must match the marking on the product. Use the supplied plastic
sealing rings to seal the connection.
Compressed air connection
Air inlet (WALL)
Connect the compressed air line from the central distribution line to the
WALL quick coupling - inlet compressed air (if a central air distribution line is
available).
Air outlet (OUT)
Connect the pressure hose to the medical ventilator to the OUT quick
coupling on the compressed air outlet.
Air from the central distribution system is automatically connected through
the compressor to the OUT outlet port. In this configuration, the compressor
serves as a backup source of compressed air. If the air pressure from the
central distribution system is reduced, the compressor automatically switches
on and there is no interruption in the supply of continuously pressurized air at
the outlet of the compressor.
The maximum rated air pressure for the WALL quick coupling is 5.5 bar
/ 80 psi. Please note that the air supply from the central distribution line
entering the compressor must be medical grade air (particulate size
and moisture content). The compressor does not modify the air from
the central distribution line.
NP-DK50 DI-17_03-2021
17
OPERATION