5. MAINTENANCE
REPAIRS AND SERVICE
Warranty and extended warranty repairs are to be completed by the
manufacturer or a service provider authorized by the manufacturer.
Manufacturer will make available on request information that will assist
service personnel to repair of medical device.
The manufacturer reserves the right to modify the equipment in any
way that will not alter the function or the operation of the equipment.
Only a qualified technician or the Customer Service Department of the
manufacturer may perform repairs that go beyond routine maintenance.
Use only spare parts and accessories approved by the manufacturer.
Prior to any maintenance or repairs, switch off the compressor and
disconnect it from the mains (pull out the main power plug).
Air pump components (head, cylinder, pressure hose, etc.) are very hot
during and shortly after compressor operation – do not touch these
components!
Allow the equipment to cool down before maintenance, service or
connecting/disconnecting the compressed air supply!
Automatic start: when pressure in the pressure tank decreases below
the switch-on pressure, the compressor automatically switches on. The
compressor automatically switches off when pressure in the air tank
reaches the switch-off pressure.
MAINTENANCE SCHEDULE
The operating entity is obliged to ensure that all tests of the equipment
are carried out repeatedly at least once within every 24 months (EN
62353) or in intervals as specified by the applicable national legal
regulations. A report must be prepared on the results of the tests
(e.g.: according to EN 62353, Annex G), including the measurement
methods used.
Maintenance
Clean air inlet filter
Preventive maintenance after 5,000 hrs (KIT DI
5K)
- replacement of microfilter
- cleaning of internal parts by compressed air
- tightness inspection
Preventive maintenance after 10,000 hrs (KIT
DI 10K)
Every 5,000 hours of operation must be carried out the preventive maintenance.
NP-DK50 DI-17_03-2021
Page
Time interval
At least once a
19
week
Maintenance
5,000 hours
documentation
Maintenance
10,000 hours
documentation
21
MAINTENANCE
To be
performed by
Staff
Qualified
technician
Qualified
technician