Inspect the unit for damages during transport. The claims for possible damages will be accepted only in case they are
immediately communicated to the carrier; the damages that are found must be written down and presented to LED SpA or to
your own retailer. If the unit is returned to the LED SpA or to your own retailer, it is necessary to use the original
equipment's package or another equivalent one, to guarantee the safety during the transport.
•
Unpack the equipment and carefully study the documentation and operating instruction supplied. Mains voltage, indicated
above the inlet, must agree with the local mains voltage (mains voltage frequency: 50-60 Hz). The correct voltage setting is
selected by turning the voltage selector when available. Insert the correct fuses in the module referring to the value written
on the label.
•
Connect mains cable to a mains outlet having good hearth connection
OPERATION OF THE EQUIPMENT WITHOUT EARTH CONNECTION IS FORBIDDEN
•
The unit must be installed on a level surface, with dimension, at least, correspondent to those of the base of the unit itself.
Around the unit must be left a space of 25cm, at least.
•
Connect the mains cable to the mains socket on the rear panel of the unit.
•
Connect the equipotential binding post located at the left of the unit's back panel to equipotential socket of the plant.
•
Connect the single foot switch or the double foot switch (optional) to the connector on the rear panel.
•
Connect handle, in the case of use of handle without finger switch it shall be connected on the black buckle.
•
In case of use of bipolar forceps (see figure paragraph 4.4.6) it is necessary to use the special adapter (REF 00498.04).
•
Let unit work in dry environment only. Any verified condenses must be let evaporate before putting in operation the unit.
Don't exceed the temperature environment or the allowed moisture. Environmental conditions: Temperature: 10/40°C -
Relative moisture: 30/75% - Pressure: 70/106k Pa
•
When the unit is switched on, through the on/off switch on the frontal panel, after having checked the internal parameters, it
will work with the function and the power level utilized during the last switching (when the unit is switched for the first time
the level will be 00).
•
Before using the unit, it is necessary connect the cable to the patient plate. Both when single plate electrodes and when split
plate electrodes are used it is necessary to confirm the impedance acceptance by pressing the key OK (see paragraph 4.3.2).
In this way, if the value of the impedance is acceptable, the OC indicator light will stop flashing and the alarm to sound.
•
Having:
Holder-handle with two pushbuttons without foot switch: press the yellow pushbutton on the holder-handle to deliver the
cutting current (the choice between CUT, BLEND and BIPOLAR CUT must be done pressing the correspondent pushbutton on
the unit); or the blue pushbutton on the holder handle to deliver coagulating current (the choice between FORCED COAG,
SOFT COAG and BIPOLAR COAG must be done pressing the correspondent pushbutton on the unit).
Holder handle with two pushbuttons and a single foot switch: choose the cutting current CUT, BLEND or BIPOLAR CUT and
the coagulation current FORCED COAG, SOFT COAG or BIPOLAR COAG. Preset through the yellow pushbutton on the
holder handle, the function for the cut that appears on the unit or, through the blue pushbutton on the holder handle, the function
for the coagulation that appears on the unit. The current delivery takes place through the foot switch.
Holder handle with two pushbuttons and double foot switch: press the yellow foot switch or the yellow pushbutton of the
holder handle to pre-set and deliver the cutting current (the choice between CUT, BLEND and BIPOLAR CUT must be done
pressing the correspondent pushbutton on the unit) or the blue foot switch or the blue pushbutton of the holder handle to pre-set
and deliver the coagulating current (the choice between FORCED COAG, SOFT COAG and BIPOLAR COAG must be done
pressing the correspondent pushbutton on the unit).
Holder handle without pushbuttons and single foot switch: connect the holder handle to the black binding post and pre-set the
current for the cut (CUT, BLEND or BIPOLAR CUT) or the coagulation (FORCED COAG, SOFT COAG or BIPOLAR
COAG), press the foot switch to deliver the pre-set current.
Holder handle without pushbuttons and double foot switch: connect the holder handle to the black binding post and press the
yellow footswitch to pre-set and deliver the cutting current (the choice between CUT, BLEND and BIPOLAR CUT must be
done pressing the correspondent pushbutton on the unit); press the blue foot switch to pre-set and deliver the coagulating current
(the choice between FORCED COAG, SOFT COAG and BIPOLAR COAG must be done pressing the correspondent
pushbutton on the unit).
Bipolar forceps and single foot switch: connect the furnished adapter (see paragraph 4.4.6). The equipment will select the
BIPOLAR operative mode (BIPOLAR CUT or BIPOLAR COAG), choose the desired function pressing the corresponding
button. To deliver the current press the foot switch. To avoid the forcep's damage don't make short circuit with its tips.
Bipolar forceps and double foot switch: connect the furnished adapter (see paragraph 4.4.6). The equipment will select the
BIPOLAR operative mode (BIPOLAR CUT or BIPOLAR COAG), choose the desired function pressing the corresponding
button. To deliver the current press the foot switch for the coagulation (blue). To avoid the forcep's damage don't make short
circuit with its tips.
Manuale d'istruzioni / Instruction's Manual
3. INSTALLATION
MA182IGBb
DIATERMO MB200
25