2.2 INSTALLATION

The electric safety is insured only when the same are correctly connected to an efficient net linked to the earth in conformity
with the actual safety requirements. It is necessary to verify this fundamental safety requisite and, in case of doubt, to
require an accurate control of the plant from part of qualified personnel. The manufacturer cannot be considered responsible
for possible damages caused from the lack of efficient connection to earth of the installation. Operation without a protective
earth connection is forbidden.
Before connect the equipment ascertain that the required voltage (showed on the rear panel) corresponds to the available
mains.
In case of incompatibility between the available wall socket and the feeding cable of the equipment, replace only with
legally approved connectors and accessory items. The use of adapters, multiple connections or cable extensions is not
advised. Should their use become necessary it is mandatory to use only simple or multiple adapter conforming to the actual
safety requirements.
Don't let the apparatus exposed to atmospheric agents. The unit must be protected from seepage of liquids.
Don't obstruct openings or cracks of ventilation or heathsink
Don't leave the equipment uselessly inserted. Switch off the equipment when not in use.
The use of the unit is not suited in explosive rooms.
must be destined only to the use for that have been expressly designed. Any other use is to be
DIATERMO MB200
considered improper and dangerous. The manufacturer can not be considered responsible for possible damages due to
improper, wrong and unreasonable uses.
It is dangerous to modify or try modifying the characteristic of the equipment.
Before effect any operation of cleaning or maintenance, disconnect the apparatus from the electric net, either unplugging it
from the mains or switching off the mains switch of the plant.
In case failure and/or bad operation of equipment switch off it. For the possible reparation address only to an authorised
service centre and ask for the use of original spare parts. The lack to follow the above requirements could risk the safety of
the equipment and can be dangerous for the user.
Do not reduce or disable the audible signal warning the activation of the generator. A functioning activation signal can
minimise or prevent patient or staff injury in the event of accidental activation.
Avoid verifying the functioning of the unit by shorting the active electrode with the reference one or the active electrode
with metallic parts.
WARNING: When the electrosurgical unit is used in operating rooms it is necessary to just use waterproof foot-switches
(
00302.00 Water-proof pedal with single switch –
REF
Manuale d'istruzioni / Instruction's Manual
00301.03 Water-proof pedal with double switches)
REF
MA182IGBb
DIATERMO MB200