4.4.4 C
S
URRENT FOR
UPERFICIAL
The modulated current (FORCED COAG) it is characterized by good property of surface coagulation behaving at the time it
probable production of eschar and partial carbonization of the tissue. The advantage of this type of coagulation resides in the
rapidity with which the effect is gotten.
4.4.5 C
D
URRENT FOR
EEP
The low voltage and low modulation current (SOFT COAG) it is suited for coagulation of deep layers of the fabric in which the
coagulation of the cellular albumin is gotten in absence of carbonization and without production of eschar. The process of
coagulation is in this case more time expensive than that of the forced coagulation.
4.4.6 B
C
IPOLAR
OAGULATION
This current is low voltage pure sine current and suited for coagulation without carbonization either monopolar or bipolar. The
use of bipolar forceps it is allowed only with this current. To allow the connection of the cable bipolar forceps it is necessary the
use of an adapter that prevents any other type of current from delivering.

4.6 SIGNALING OF EXCESSIVE TIME OF DELIVERY (OVT)

If the operator exceeds the maximum time of disbursement, recommended by the international norms, that is 10 seconds, the
equipment produces a signal of warning consisting of al bright intermittent signal OVT. If despite the signal of warning the
operator insists in the continuous delivering, after a time depending from the type of current, and from the level of the same one,
the signal of warning is transformed in impediment to the delivering of current that is signalled through the signal OVT
constantly illuminated. The interdiction lasting of the current delivery depends from the previous conditions of delivery.
4.7 SIGNALING OF EXCESSIVE IMPEDANCE IN THE CIRCUIT OF NEUTRAL ELECTRODE
(OC)
For the meaning of this warning signal please refer to the previous description of the neutral electrode circuit.

4.8 PRESETTABLE SETTING BY USER

User is allowed to preset the following settings: Acoustic signals level; power level steps.
To modify the acoustic signals level do as for the following instructions::
Switch on the unit while the CUT pushbutton is maintained pressed.
When the unit has finished the internal parameters check, on the CUT display will appear the message SOU. and on the COAG
display will appear the preset value.
Release the CUT pushbutton. Set the new acoustic level by the COAG pushbuttons. While the level adjustment the
corresponding sound level can be heard.
Manuale d'istruzioni / Instruction's Manual
C
(FORCED COAG)
OAGULATION
C
(SOFT COAG)
OAGULATION
C
(BIPOLAR COAG)
URRENT
Sound emission at 1m distance
Level
from the frontal panel
1
55 dBA
2
60 dBA
3
65 dBA
4
70 dBA
5
75 dBA