Masimo Rad-67 Mode D'emploi page 3

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 5
Data Transfer Download Cable
Rad-67™
Reusable
Prior to using this cable, the user should read and understand the Operator's Manual for the Device, and Software, the Directions for
Use for the Sensor, and this Directions for Use.
INTENDED USE
The Masimo® Rad-67™ data transfer download cable is intended for use by trained personnel to download patient data
from Rad-67, such as functional saturation of arterial oxygen hemoglobin (SpO
saturation (SpCO®), methemoglobin saturation (SpMet®), and total hemoglobin concentration (SpHb®) in clinical and non-
clinical settings.
DESCRIPTION
The Rad-67 data transfer download cable ("Cable") is an accessory for Rad-67 ("Device"). The cable connects to the device via
power charging connector (Masimo proprietary port) and transfers data (download only) to computer storage via USB connector.
The Rad-67 data transfer download cable can also be used for device software upgrades via USB software upgrade tool.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed to
use Masimo sensors.
WARNINGS, CAUTIONS AND NOTES
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• Masimo sensors and cables are designed for use with devices containing Masimo SET oximetry or licensed to use
Masimo sensors.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
INSTRUCTIONS
NOTE: Patient monitoring must be suspended while downloading trend data from the device.
1. Consult your Masimo Representative for instructions on installing and operating data download software.
2. Refer to Fig. 1. Review the connections required for data output.
3. Refer to Fig. 2. Insert the Masimo Proprietary cable end into the Masimo Proprietary Port on the device.
4. Refer to Fig. 3. Connect the female USB end of the cable to a USB Null Modem Cable (not included).
5. Refer to Fig. 4. Insert the USB Null Modem Cable into the USB port on the PC.
6. Ensure all cables are securely connected.
7. Download data.
8. Refer to Fig. 5. Pull gently to disconnect the Data Transfer Download Cable from the device.
CLEANING
1. Clean the Data Transfer Download Cable by thoroughly wiping it with 70% isopropyl alcohol, glutaraldehyde, 10%
bleach (sodium hypochlorite) to 90% water solution, or quaternary ammonium chloride solution.
2. Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.
3. Allow the cable to dry completely before storage or next use.
CAUTION: Do not soak or immerse the cable in any liquid solution. Do not attempt to sterilize by irradiation, steam or
ethylene oxide.
ENVIRONMENTAL
Storage Temperature
Relative Humidity
COMPATIBILITY
Masimo Cables are for use only with devices containing Masimo SET® and/or Masimo rainbow® technology
and with compatible sensors. Consult individual oximetry system manufacturer for compatibility of particular
devices and sensor models. Each device manufacturer is responsible for determining whether their devices are
compatible with each sensor and/or cable model.
For Compatibility Information Reference: www.Masimo.com
D I R E C T I O N S F O R U S E
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
-4°F to 158°F (-20°C to 70°C) @ ambient humidity
10% to 95% RH (non-condensing) @ ambient temperature
), pulse rate (PR), carboxyhemoglobin
2
3
en
Non-sterile
9871B-eIFU-0319

Publicité

Table des Matières
loading

Table des Matières