Operating Instructions
8.3: Analogue output behavior in the event of an error
I / mA
20,5
20,0
4,0
3,8
I / mA
20,5
20,0
4,0
3,8
UL
Lower limit of measuring range
Figure 15: Analogue output behavior in the event of an error (1)
Valid range of current output according to NAMUR NE43: Max. current output 20.5 mA / Min. current output 3.8 mA.
There might be a discontinuity between the linear range MBA...MBE...OL and the range between MBE and OL. The same for UL...MBA.
7 6
O p e r a t i n g I n s t r u c t i o n s | S I C K
8 Errors
Analog output
4 mA ... 20 mA
(normal function)
Analog output
4 mA ... 20 mA
(inverted mode)
 20 mA ... 4 mA
OL
Upper limit of measuring range
p
p
8017612/YSZ4/2016-02-09 | PAC50