Signal Description - Agilent Technologies TV 401/301 NSF Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41

Signal Description

START/STOP: input signal to start or stop the pump. With the
supplied cover connector the START/STOP (+) signal is connected to
the +24 Vdc pin and the START/STOP (-) signal to the GROUND pin:
in this condition the pump automatically starts as soon as the
controller recognises the input supply ("Plug & Pump").
INTERLOCK: input signal to control the pump rotation. With the
supplied cover connector the INTERLOCK (+) signal is connected to
the +24 Vdc pin and the INTERLOCK (-)signal to the GROUND pin.
Speed setting: PWM input signal to set the pump speed. The PWM
signal characteristics must be the following:
Figure 16
If any signal isn't applied (no connection) the pump speed is set to
"High Speed" (1010 Hz max and settable by window "120").
TV 401/301 NSF User Manual / 87-900-963-01
frequency: 100 Hz +/-20 %
amplitude: 24 V max
duty cycle range: from 25 % to 75 %
corresponding to an output frequency from 750 Hz to 1010 Hz
(see the following diagram):
16BTechnical Information
148BInterconnections
(E)
17
253/290

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-8928

Table des Matières