1 Automatic means that when the controller stops, the vent valve is
NOTE
opened with a delay defined by window n. 126; when the controller
starts, the vent valve is immediately closed.
2 On command means that the vent valve is opened or closed by
means of window n. 122.
3 These commands remove the pump motor supply for a short time.
If they are used continuously, they can cause the pump stopping.
TV 401/301 NSF User Manual / 87-900-963-01
149BRS 232/RS 485 Communication Description
Read/Write
Datatype
R
N
Reserved to Agilent service
R
A
R
A
R
A
Reserved to Agilent service
R/W
N
R/W
L
16BTechnical Information
Description
Pump life in hours (zeroed by
the reset command)
CRC EPROM (QE)
CRC Param. (PA)
CRC Parameter structure
RS 485 address
Serial type select
(E)
17
Admitted Values
999999
QE8XXXX
(where "XXXX"
are variable)
PA8XXXX
(where "XXXX"
are variable)
"XXXX"
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)