TV 401/301 NSF User Manual / 87-900-963-01
149BRS 232/RS 485 Communication Description
Read/Write
Datatype
R/W
L
R/W
L
R/W
N
W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
R/W
L
16BTechnical Information
Description
Water cooling
Active Stop
(write only in stop)
Baud rate
Pump life/ cycle time/ cycle
number reset
Interlock type
(default = 1)
Analog output type: output
voltage signal proportional to
frequency or power
Low Speed frequency (Hz)
Rotational frequency setting
(Hz)
Maximum rotational
frequency in Hz (active only
in Stop condition)
Set vent valve on/off
(on = closed)
(E)
17
Admitted Values
0 = NO
1 = YES
0 = NO
1 = YES
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
750 to "Maximum
rotationl
frequency"
(win 121)
(default = 750)
750 to "Maximum
rotationl
frequency"
(win 121)
(default = 1010)
750 to 1010
(default = 1010)
On = 1
Off = 0
(default = 1)
269/290