Télécharger Imprimer la page

Piezon ® Handpiece And - EMS AIRFLOW Prophylaxis Master Mode D'emploi

Masquer les pouces Voir aussi pour AIRFLOW Prophylaxis Master:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

®
2.6. PIEZON
Handpiece and Instruments
®
PIEZON
instruments and handpieces are reusable, but they shall be reprocessed before use:
cleaned, disinfected and sterilized. Non-sterile handpieces and accessories may cause bacterial or
viral infections.
Follow the "Reprocessing of EMS parts" instructions and the present-day regulations on
reprocessing enforced in your country.
Check tip length and tip thread through the
cover right folder of your Quick Guide.
If tip extremity is in the red area, it can have
excessive and uncontrolled vibration. Replace the
tip.
1
Mount the tip / insert using the
EMS CombiTorque tool
Once the instrument is screwed
all the way in, give an extra quarter of
a turn to obtain the required torque
and remove the CombiTorque.
Use only the CombiTorque to
tighten the EMS instrument on the
handpiece to the correct torque to
avoid tip unscrewing or breaking.
ENGLISH
2
Connect the PIEZON
handpiece
Blow-dry the connections to
remove any presence of liquid and to
ensure a proper electrical contact.
Never blow compressed air in the
irrigation connections, as this will
irremediably damage internal parts.
FB-618/EN – Rev. A-01 - 2017.03.30
Accessories are available from
EMS and authorized dealers.
®
Nose cap and light guide
Always use the handpiece with
the light guide and the nose cap
installed and correctly tightened.
For replacing the light guide, refer to
®
the "PIEZON
replace" section.
3
light guide check &
16/46

Publicité