5.3. Reprocessing of EMS parts
EMS recommends the use of cleaning, disinfection, packaging for sterilization and sterilization
procedures accordingly with ISO 17664.
Report adverse events related to device reprocessing directly to EMS.
Reusable products must be cleaned and sterilized prior to first use. Do not reprocess the
products over the allowed number of sterilization cycles, but replace: refer to the "Service life"
section of the "Technical Description" chapter.
Concentrations and contact times specified by the manufacturer of the cleaning and
disinfection agent must be followed. Only freshly prepared solutions shall be used.
Remember that sterilization cannot be achieved unless the elements of the assembly are
cleaned and disinfected first.
If there is anything in the following instructions that is not clear or seem to be inadequate, do not
hesitate to contact/inform EMS.