Power cannot be set

HF-cable is not connected.
No radiator is connected.
No treatment time is set.
9.2.4
During treatment the HF-power switches off and a sound signal is audible (3 beeps of 1
sec.)
The HF-cable is mistakenly disconnected from radiator or apparatus.
The HF-cable is damaged and must be replaced.
Bad contact occurred in the connection of the HF-cable to the apparatus and/or radiator.
9.3

Technical maintenance

We recommend to have the unit checked annually. This may be done by your supplier, or by another
agency, authorised by Enraf-Nonius. It is also recommended that a record of the service history is kept for
all activities relating to service and maintenance. In some countries this is even obligatory.
Maintenance and all repairs should only be carried out by an authorised agency. Enraf-Nonius will not be
held responsible for the results of maintenance or repairs by unauthorised persons.
Opening of the equipment by unauthorised agencies is not allowed and will terminate any claim to
warranty.
9.4

Expected Service Life

Expected service life of this device is 8000 hours of use or 10 years after purchase, whichever comes
first.
9.5

End of life

Your Radarmed and accessories contains materials which can be recycled and/or are
noxious for the environment. At the end of the duration of life, specialised concerns can
take apart these articles and sort out the noxious materials and materials for recycling.
By doing so you contribute to a better environment.
Please ensure that you are well informed of the local rules and regulations
regarding to the removal of apparatus and accessories.

10 Specifications

10.1 Microwave therapy

Frequency
HF power, continuous
HF power, pulsed
Operating modes
Clock

10.2 Technical specifications

Mains voltage
Frequency
Current consumption
Safety class
Medical device classification
Weight
Dimensions
081-400-095-49_EN
: 2450 MHz +/- 50 MHz
: 0 - 250 Watt +/- 20% at 50 Ohm
: 0 - 250 Watt +/- 20% with a fixed peak power of 1500 Watt
: continuous and pulsed
: 0 - 30 minutes
: 220-240 Volt
: 50 / 60 Hz
: 1100 VA
: 1 type B
: IIa (according to Medical Device Directive (93/42/EEC))
: 45 kg (including arm)
: 93x39x49 cm (hxwxd including castors)