The pump must be always water cooled. For this purpose the
customer must use the dedicated channel on the pump body.
During normal operation, the motor is fed with a 60 Vac and 530 Hz
three-phase voltage. The power provided to the motor is limited by
the controller according to:
WARNING!
The user must set the gas load (win.157) before starting the pump.
For details see the following figure.
The applied power limit is readable by serial line (window 155).
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01 (D)
gas load (selectable by window 157: see the following "Window
Meanings" paragraph)
body temperature (can be read by window 216).
Technical Information
Description of the Turbo-V 3K-G
17