16
Instructions for Use
The turbopump with ISO F inlet flange must be fixed to the vacuum
chamber by means of bolts with a strength class of 12,9
(σr = 1200 N/mm
of steel bolts and the relevant fixing torque as recommended by
Agilent.
Tab. 1
FLANGE
ISO 250 F
WARNING!
The Turbo-V 3K-G cannot be fixed by means of its base. The system can only
be fixed through its ISO 250F flange.
WARNING!
Failure to comply with these installation instructions could result in the pump
detaching from the system in the event of a rotor failure, which could cause
property damage or serious injury or death.
290/360
2
). The following table shows the necessary number
FIXING DEVICE
M10 bolts
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01 (D)
N.
FIXING TORQUE
12
30 Nm