Technical Data

15 . 1 0 Metrological Data

15 . 1 0 . 1 Models BCE6202 | BCE4202 | BCE3202
Model
Readability | Scale interval (d)
Maximum capacity (Max)
Repeatability
At 5%, typical value
At approx . the maximum load, typical value
Linearity deviation
Limits
Typical value
Sensitivity drift between +10°C and +30°C
Tare maximum capacity (subtractive)
isoCAL (only for I-1x models):
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weighing according to USP (United States
Pharmacopeia), Chap . 41
Optimum minimum initial weighing
Typical measurement time
Typical stabilization time
Recommended Calibration Weight
External calibrated test weight
Accuracy class in accordance with OIML R111-1
Weighing pan size
Net weight, approx .
42
Entris
BCE Operating Instructions
®
Unit
Value
Value
mg
10
10
g
6200
4200
mg
5
5
mg
10
10
± mg
20
20
± mg
6
6
± ppm/K
2
2
<100% of maximum capacity
K
2
2
h
6
6
II
II
BC-EE
BC-EE
mg
100
100
mg
500
500
g
8 .2
8 .2
s
≤1.0
≤1.0
s
≤0.9
≤0.9
g
5000
2000
F1
F1
mm
182 x 182
182 x 182
kg
4 .6 | 6 .2
4 .6 | 6 .2
Value
10
3200
5
10
20
6
2
2
6
II
BC-EE
100
500
8 .2
≤1.0
≤0.9
2000
F1
182 x 182
4 .6 | 6 .2

Publicité

Table des Matières
loading

Table des Matières