System Settings
7 .3 .5

"DEVICE"/"EXTRAS" Menu

Parameter
Setting values
MENU
EDITABL .*
RD .ONLY
SIGNAL
OFF
ON*
KEYS
UNLOCK .*
LOCKED
EXT .KEY
PRINT
CAL .
CF
ENTER
ZERO
TARE
GLP END
NO*
ON/STDB .*
ON MODE
ON/OFF
AUTO ON
OFF
LIGHT
ON*
* Factory setting
7 .3 .6

"DATA .OUT ."/"COM . SBI" Menu

Parameter
Setting values
COM . OUTP . IND .NO*
IND .AFTR
AUTO . W /O
AUTO W/
* Factory setting
22
Entris
BCE Operating Instructions
®
Explanation
Activates write access . The menu parameters can be changed .
Activates read access . The menu parameters cannot be changed .
Switches the acoustic signal off .
Switches the acoustic signal on .
Deactivates the button lock .
Activates the button lock .
Assigns the print function to the external key .
Assigns the calibration and adjustment function selected under [CAL ./ADJ .] to
the external key .
Assigns the [Back] function to the external key .
Assigns the [Confirm] function to the external key .
Assigns the [Zero] function to the external key .
Assigns the [Tare] function to the external key .
Assigns the [GLP] function to the external key .
No function is assigned to the external key .
The [On/Off] button switches between on/standby with time .
The [On/Off] button switches between on/standby without time . .
Changes the function of the [On/Off] button: The device no longer switches off
or to standby mode, instead it starts a boot process .
Deactivates the lighting on the operating display .
Activates the lighting on the operating display .
Explanations
Activates the manual data output without stability .
Activates the manual data output after stability .
Activates the automatic data output without stability .
Activates the automatic data output after stability .