Technical Data
15.10 Metrological Data
15.10.1 BCE224 | BCE124 | BCE64 | BCE653 | BCE623 | BCE423 | BCE323 | BCE223 models
Model
Readability | Scale interval (d)
Maximum capacity (Max)
Repeatability
At 5% load, typical value
At approx . maximum load, typical value
Linearity deviation
Limits
Typical value
Sensitivity drift between +10 °C and +30 °C
Tare maximum capacity (subtractive)
isoCAL (only for i-1x models):
Temperature change
Time interval
For models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap . 41
Optimum minimum initial weighing
Typical measurement time
Typical stabilization time
Recommended calibration weight
External calibrated test weight
Accuracy class in accordance with
OIML R111-1
Weighing pan size
Weighing chamber height (upper edge of the
weighing pan to the lower edge of the upper
draft shield panel)
Net weight, approx .
40
Entris
®
BCE Operating Instructions
Unit
Value
Value
Value
mg
0 .1
0 .1
0 .1
g
220
120
60
mg
0 .08
0 .08
0 .08
mg
0 .1
0 .1
0 .1
± mg
0 .2
0 .2
0 .2
± mg
0 .06
0 .06
0 .06
± ppm/K 1 .5
1 .5
1 .5
<100% of maximum capacity
K
1 .5
1 .5
1 .5
h
4
4
4
I
I
I
BC_EB BC_EB BC_EB BC_ED BC_ED BC_ED BC_ED BC_ED
mg
1
1
1
mg
10
10
10
g
0 .082
0 .082
0 .082
s
<2 .0
<2 .0
<2 .0
s
<1 .5
<1 .5
<1 .5
g
200
100
50
E2
E2
E2
mm
Ø 90
Ø 90
Ø 90
mm
240
240
240
kg
5 .9 |
5 .9 |
5 .9 |
6 .2
6 .2
6 .2
Value
Value
Value
Value
1
1
1
1
650
620
420
320
0 .5
0 .5
0 .5
0 .5
1
1
1
1
2
2
2
2
0 .6
0 .6
0 .6
0 .6
2
2
2
2
2
2
2
2
6
6
6
6
II
II
II
II
10
10
10
10
20
20
20
20
0 .82
0 .82
0 .82
0 .82
< 1 .5
<1 .5
<1 .5
<1 .5
<1 .0
<1 .0
<1 .0
<1 .0
500
500
200
200
F1
F1
F1
F1
Ø 120 Ø 120 Ø 120 Ø 120 Ø 120
50
240
240
240
4 .5 |
5 .9 |
5 .9 |
5 .9 |
4 .8
6 .3
6 .3
6 .3
Value
1
220
0 .5
1
2
0 .6
2
2
6
II
10
20
0 .82
<1 .5
<1 .0
200
F1
240
5 .9 |
6 .3