14 . 1 2 Metrological Data

14 . 1 2 . 1 Models BCA324 | BCA224 | BCA124 | BCA64

Model
Readability | Scale interval (d)
Maximum capacity (Max)
Repeatability
At 5%, typical value
At approx . the maximum load, typical value
Linearity deviation
Limits
Typical value
Sensitivity drift between +10°C and +30°C
Tare maximum capacity (subtractive)
isoCAL (only for I-1x models):
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weight according to USP (Unit-
ed States Pharmacopeia), Chap . 41
Optimum minimum initial weight
Typical stabilization time
Typical measurement time
Typical stabilization time
Recommended calibration weight
External calibrated test weight
Accuracy class in accordance with
OIML R111-1
Unit
mg
0 . 1
0 . 1
g
320
220
mg
0 .08
0 .08
mg
0 . 1
0 . 1
± mg
0 .3
0 .2
± mg
0 .06
0 .06
±
1 .0
1 .0
ppm/K
<100% of maximum capacity
K
1 .5
1 .5
h
4
4
I
I
BC-AA
BC-AB
mg
1
1
mg
10
10
g
0 .082
0 .082
0 . 1 6
0 . 1 6
s
≤2.0
≤2.0
s
≤1.5
≤1.5
g
200
200
E2
E2
Entris
II Advanced Line Operating Instructions
®
Technical Data
0 . 1
0 . 1
120
60
0 .08
0 .08
0 . 1
0 . 1
0 .2
0 .2
0 .06
0 .06
1 .0
1 .0
1 .5
1 .5
4
4
I
I
BC-AB
BC-AB
1
1
10
10
0 .082
0 .082
0 . 1 6
0 . 1 6
≤2.0
≤2.0
≤1.5
≤1.5
100
50
E2
E2