Single Point Calibration; Dual Point Calibration - Mettler Toledo InPro Serie Instructions D'utilisation

Masquer les pouces Voir aussi pour InPro Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

InPro 6850 i / 6900 i / 6950 i O
–  For correct calibration, a minimum flow rate of 
the calibration medium is necessary.
–  Make sure that all other parameters, such as 
  t emperature and pressure, are constant.
For continuous applications, we recommend peri-
odic recalibration in line with your requirements
on accuracy, the type of process in operation and
your own experience. The frequency of the need 
for re-  c alibration depends very much on the specific 
  a pplication, and therefore appropriate intervals can-
not be exactly defined here.

5.2.3 Single point calibration

By carrying out a single point calibration, the factual 
slope of the sensor can be established. The cali-
bration medium can be water with known oxygen 
saturation index (e.g. air-saturated water) or air with 
known   w ater-vapor saturation (e.g. water-vapor 
saturated air).
After the sensor signal has stabilized, the complete 
meas  u ring system can then be calibrated to the 
100  % value of the   d esired measurable variable, 
e.g.100 % air, 20.95 % O
(77 °F) and normal pressure (see instruction manual 
for the   t ransmitter).

5.2.4 Dual point calibration

By carrying out a dual point calibration both slope 
and zero point of the sensor can be established. 
a
Important! In case of a dual point calibration,
always start by the zero point calibration before
calibrating the slope.
As a rule, the zero point should be adjusted to zero 
manually, or it is automatically performed by the 
transmitter (see instruction manual for the transmit-
ter).
Zero-point calibration is recommended:
–  when measuring less than 5 ppb in liquids
–  when measuring less than 125 ppm (by vol.) in 
gases
a
Attention! Incorrect zero point calibration is a
frequent source of measurement error. For cor-
rect calibration, we recommend the use of car-
bon dioxide gas with a level of purity of at least
99.9995 %.
After the sensor signal has stabilized (after 6 ... 12
hours), the sensor can be calibrated through the 
  r elevant transmitter to the zero % value of the desired 
measurable variable, e.g. 0 % air, 0.0 % O
0.0  ppm (see instruction manual for the transmitter).
© 04 / 13 Mettler-Toledo AG   
Printed in Switzerland 
 Sensor 12 / 25 mm  
2
, or 8.26 ppm at 25 °C 
2
InPro 6850 i / 6900 i / 6950 i 
19
, or 
2
52 206 349

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Inpro 6850iInpro 6900iInpro 6950i

Table des Matières