Changing settings during the centrifugation run.

The running time, the speed, the relative centrifugal force (RCF/RZB), the run-up and run-down parameters as well
as the temperature can be changed while the centrifuge is running.
The parameters can only be changed one at a time and in succession.

Select the desired parameter and change the value using the control knob
centrifugation parameters").

Press the
key. The modified setting will be stored in program location "----" (see Chapter "Automatic
START
buffer"). The original program will not be overwritten.
18
Integral RCF (∫RCF)
The integral RCF (∫RCF) is a measure for the sedimentation effect (∫ n
centrifugation runs. To interrogate the integral RCF, press and hold the
19

Displaying the maximum speed of the rotor


Press the
key repeatedly until the parameter RPM is displayed and the input field is shown inversely.
n

Press the
key once more and hold it pressed; the maximum speed of the rotor (n-max-Rotor) will be displayed.
n
20

Displaying the maximum RCF of the rotor


Press the
key repeatedly until the parameter RCF/RZB is displayed and the input field is shown inversely.
RCF

Press the
key once more and hold it pressed; the maximum RCF of the rotor (RCF-max-Rotor) will be
RCF
displayed.
21

Emergency stop


Press the
key twice. The STOP symbol will flash in the display.
STOP
In the case of an emergency stop, run-down takes place with braking step R9 (shortest run-down time). If braking
step R0 was selected, the run-down time will be longer than with braking step R9 for technical reasons.
22

Audible signal

The audible signal sounds in accordance with the following philosophy:
OFF

at 2 s intervals on the occurrence of a fault
ON1

at 2 s intervals on the occurrence of a fault

at 30 s intervals on completion of a centrifugation run and the rotor coming to rest.

ON2
at 2 s intervals on the occurrence of a fault

at 30 s intervals on completion of a centrifugation run and the rotor coming to rest.

every time a key is pressed
The audible signal is silenced by opening the lid or pressing any key.
The signal can be set as follows when the rotor is at standstill.

Open the lid.

Press and hold the
key for 8 seconds.
t
After 8 seconds SOUND / BELL XXX will be displayed.

Set the desired function (OFF, ON1, ON2) by means of the control knob

Press the
key to save the setting.
START
 ok  will be displayed briefly as a confirmation.
23

Interrogating the operating hours

It is only possible to interrogate the operating hours when the rotor is at standstill.

Open the lid.

Press and hold the
key for 8 seconds.
t
After 8 seconds SOUND / BELL XXX will be displayed.

Press the
key once more.
t
The operating hours (CONTROL: ) for the centrifuge will be displayed.
The operating hours display will go off automatically after 10 seconds.
50/137
(see Chapter "Entering
2
dt). This numerical value is used to compare
key.