Changing settings during the centrifugation run.
The running time, the speed, the relative centrifugal force (RCF/RZB), the run-up and run-down parameters as well
as the temperature can be changed while the centrifuge is running.
The parameters can only be changed one at a time and in succession.
Select the desired parameter and change the value using the control knob
centrifugation parameters").
Press the
key. The modified setting will be stored in program location "----" (see Chapter "Automatic
START
buffer"). The original program will not be overwritten.
18
Integral RCF (∫RCF)
The integral RCF (∫RCF) is a measure for the sedimentation effect (∫ n
centrifugation runs. To interrogate the integral RCF, press and hold the
19
Displaying the maximum speed of the rotor
Press the
key repeatedly until the parameter RPM is displayed and the input field is shown inversely.
n
Press the
key once more and hold it pressed; the maximum speed of the rotor (n-max-Rotor) will be displayed.
n
20
Displaying the maximum RCF of the rotor
Press the
key repeatedly until the parameter RCF/RZB is displayed and the input field is shown inversely.
RCF
Press the
key once more and hold it pressed; the maximum RCF of the rotor (RCF-max-Rotor) will be
RCF
displayed.
21
Emergency stop
Press the
key twice. The STOP symbol will flash in the display.
STOP
In the case of an emergency stop, run-down takes place with braking step R9 (shortest run-down time). If braking
step R0 was selected, the run-down time will be longer than with braking step R9 for technical reasons.
22
Audible signal
The audible signal sounds in accordance with the following philosophy:
OFF
at 2 s intervals on the occurrence of a fault
ON1
at 2 s intervals on the occurrence of a fault
at 30 s intervals on completion of a centrifugation run and the rotor coming to rest.
ON2
at 2 s intervals on the occurrence of a fault
at 30 s intervals on completion of a centrifugation run and the rotor coming to rest.
every time a key is pressed
The audible signal is silenced by opening the lid or pressing any key.
The signal can be set as follows when the rotor is at standstill.
Open the lid.
Press and hold the
key for 8 seconds.
t
After 8 seconds SOUND / BELL XXX will be displayed.
Set the desired function (OFF, ON1, ON2) by means of the control knob
Press the
key to save the setting.
START
ok will be displayed briefly as a confirmation.
23
Interrogating the operating hours
It is only possible to interrogate the operating hours when the rotor is at standstill.
Open the lid.
Press and hold the
key for 8 seconds.
t
After 8 seconds SOUND / BELL XXX will be displayed.
Press the
key once more.
t
The operating hours (CONTROL: ) for the centrifuge will be displayed.
The operating hours display will go off automatically after 10 seconds.
50/137
(see Chapter "Entering
2
dt). This numerical value is used to compare
key.
RCF
.