EN
3
Technical specifications
Manufacturer
Model
Type
Mains voltage ( 10%)
Mains frequency
Connected load
Current consumption
Cooling medium
Max. capacity
Allowed density
Speed (RPM)
Force (RCF)
Kinetic energy
Obligatory inspection (BGR 500)
Ambient conditions (EN / IEC 61010-1)
Set-up site
Altitude
Ambient temperature
Humidity
Excess-voltage category
(IEC 60364-4-443)
Pollution degree
Device protection class
EMC
Emitted interference,
Interference immunity
Noise level (dependent on rotor)
Dimensions
Width
Depth
Height
Weight
38/137
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
ROTO SILENTA 630 RS
5005,
5005-20,
5005-80
5005-50,
SK 48.06-30
9700 VA
9400 VA
14 A
13.5 A
R 404A
Up to 2000 m above sea level
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Not suitable for use in explosion-endangered areas.
EN / IEC 61326-1, Class B
62 dB(A)
1015 mm
approx. 355 kg
approx. 367 kg
5005-90
400 V 3~ +N
50 - 60 Hz
6600 VA
9.5 A
12000 ml
3
1.2 kg/dm
6000
6520
215000 Nm
Yes
Indoors only
5°C to 40°C
2
56 dB(A)
813 mm
1050 mm
973 mm
approx. 306 kg
approx. 303 kg
5005-60
----