13. EMC / Technical description
The Swing breast pump needs special precautions regarding EMC (Electromagnetic
compatibility) and needs to be installed and put into service according to the EMC
information provided in these instructions for use. Wireless communications equip-
ment such as wireless home network devices, mobile phones, cordless telephones
and their base stations, walkie-talkies can affect the electric breast pump Swing and
should be kept at least a distance 1.0 m away from the equipment.
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 1)
Electromagnetic emissions
The electric breast pump Swing is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breast pump Swing should assure
that it is used in such an environment.
Emission tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
a
No significant voltage changes
Warning – The electric breast pump Swing should not be used adjacent
160
to or stacked with other equipment. If adjacent or stacked use is necessary,
the electric breast pump Swing should be observed to verify normal operation in
the configuration in which it will be used.
Electromagnetic environment –
Compliance
guidance
The electric breast pump Swing uses RF en-
ergy only for its internal function. Therefore, its
Group 1
RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
Class B
The electric breast pump Swing is suitable for
use in all establishments, including domestic
Class A
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic pur-
Not
poses.
applicable