13. EMC / Technical description
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the meas-
ured field strength in the location in which the electric breast pump Swing is used exceeds the
applicable RF compliance level above, the electric breast pump Swing should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the electric breast pump Swing.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.