ERROR MESSAGES
ERROR MESSAGES
ERROR MESSAGES
ERROR MESSAGES
ERROR MESSAGES
OVERRANGE, UNDERRANGE AND BURNOUT
OVERRANGE, UNDERRANGE AND BURNOUT
OVERRANGE, UNDERRANGE AND BURNOUT
OVERRANGE, UNDERRANGE AND BURNOUT
OVERRANGE, UNDERRANGE AND BURNOUT
INDICATIONS
INDICATIONS
INDICATIONS
INDICATIONS
INDICATIONS
The instrument shows the OVERRANGE
condition with the following indication on the
upper display:
The instrument shows the UNDERRANGE
condition with the following indication on the
upper display:
The sensor leads break can be signalled as:
- for TC input : OVERRANGE or
UNDERRANGE selected by a
solder jumper (see Fig. 12).
- for RTD input : OVERRANGE
Sensor leads short circuit detection:
On RTD input, a special test is provided to signal
OVERRANGE when input resistance is less than
15 ohm (Short circuit sensor detection).
NOTE
NOTE: When:
NOTE
NOTE
NOTE
- The instrument is set for one output only and an
OVERRANGE is detected, the OUT 1 turns
OFF (if reverse action) or ON (if direct action).
- The instrument is set for heating/cooling action
and an OVERRANGE is detected, OUT 1 turns
OFF and OUT 2 turn ON.
- The instrument is set for one output only and an
UNDERRANGE is detected, the OUT 1 turns
ON (if reverse action) or OFF (if direct action).
- The instrument is set for heating/cooling action
and an UNDERRANGE is detected, OUT 1
turns ON and OUT 2 turns OFF.
ERRORS
ERRORS
ERRORS
ERRORS
ERRORS
Diagnostics are made at instrument switch-on
and during normal mode of operation.
If a fault condition (error) is detected, the lower
display will show the message "Er" while the
upper display shows the relative error code.
ERROR LIST
ERROR LIST
ERROR LIST
ERROR LIST
ERROR LIST
100
Write EEPROM error.
150
General hardware error on the CPU card.
200
Tentative to write on protected memory.
201 - 2xx Configuration parameter error. The two
less significant digit's shown the number
of the wrong parameter (ex. 209 Err show
an Error on P9 parameter)
301
RTD input calibration data error
305
TC input calibration data error
307
RJ calibration data error
310
Current transformer input calibration data
error
400
Control parameters error
500
Auto-zero error.
502
RJ error.
510
Error during calibration procedure.
GB
GB
GB
GB
GB
19