Detaching the lift from the rail system
Lower the lift so that it stands solidly on a
surface. The lift will stop automatically when
it reaches the surface. It may be useful to use
Human Care's cart (Art.no. 50346). Feed more
belt out by pulling gently on the belt that is
active while at the same time pressing the
Down button on the hand control or control
panel. When the belt is slack it can be easily
Using the lift
Important! If the buttons on the
o
hand control or control panel are
used to feed the belt back into the
lift, the belt must be kept taut until the
entire belt has been fed in.
Press the Up button on the hand control or
control panel on the lift to lift the patient.
Press the Down button on the hand control or
control panel to lower the patient. Pressing the
button activates a soft start, and releasing it
causes a soft stop.
The direction of the lever (10) determines which
belt is engaged and which is active. If the lever
(10) is pointing towards the leg hook (5) the belt
nearest the leg hook (3) is active. If the lever (10)
is pointing towards the telescopic bar (4) the
belt nearest the telescopic bar (2) is active.
2
4
30
unhooked from the distance strap by lifting the
belt loop upwards.
Hold the active belt and turn the lever (10) to
the opposite position to release the pulled-out
belt. The belt will automatically be retracted.
Make sure the belt is kept taut and gets rolled
up in the lift straight, without folding.
Important! Do not turn the lever
o
unless both belts are connected to
their respective rail rollers. The lever (10)
must always point towards the belt that
the lift is suspended from or is to be
transferred over to! When the lift is
suspended from only one of the lift belts
and the lever (10) is turned over to the
free belt, the loaded belt will be released
if you physically raise the lift. This can
result in the lift falling, which may be
dangerous. Turn the lever (10) back
immediately to the loaded belt to avoid
unintentional disconnection!
Set the telescopic bar (4) to the prescribed
position. Ensure that the bar hooks are in the
same position on both sides. Place the patient
in the prescribed sling according to the sling
instructions and hang the sling in the hooks
on the telescopic bar and in the leg hook (5).
Check that the sling's lifting straps are properly
anchored in all hooks.